Liquid Level Inst, The instrument always reads 0% or empty

Article ID: 000005
Rating: 3.6 / 5.0 (7 votes)
Views: 10023

1) Verify liquid is present in the vessel being monitored.

2) Verify the sensor is connected to the oscillator and that the inner conductor of the connecting cables is not an OPEN circuit. Also verify that that the outer conductor of the connecting cables is not an OPEN circuit.

3) Refer to Tech Note 2.

4) If a solution cannot be found, contact an Authorized AMI Technical Support Representative.