Model 601 Features
The AMI Model 601 Energy Absorber is designed to provide a compact, fast-rampdown option of up to 5 VDC for AMI power supply systems. The Model 601 is also designed to operate safely in the event of loss of facility AC power by drawing the necessary power for internal cooling from the superconducting magnet. Upon loss of AC power to a power supply system, the Model 601 will ramp the system to zero current at a discharge rate of approximately 5 VDC.
If you have purchased multiple Model 601 units with the Parallel Operation Option, then up to five Model 601 units may be operated in parallel with a total current rating of 90% of the sum of the ratings for each individual unit. AMI can provide system solutions for up to 600 A with minimal modifications.
Specifications AC Input Ratings: 100-120 or 200-240 VAC ±10%, 50-60 Hz,
100 VA maxDC Input Ratings: 5 VDC ± 5%, 4 A max Maximum Magnet Operating Current: 130 A @ 25°C, derated linearly to
100 A @ 40°CNominal Discharge Voltage: 5.0 VDC ± 2% Discharge Voltage Temperature Coefficient: ± 150 ppm/°C max. Internal Series Resistance: 2 mW ± 1 mW Rated Operating Temperature: 0 to 40°C @ 100 A,
0 to 25°C @ 130 ARated Operating Relative Humidity: 0 to 95% noncondensing Dimensions: 3.4" H x 19" W x 19.25" D
(Allow additional 1.25" front clearance for handles)Weight: 19.3 lbs.
Rev. 4 Model 601 manual in Adobe Acrobat 3.0 format (model601.pdf, 135 KB)
Back to Superconducting Magnet
Systems page
AMI HomePage |
Sales | Superconducting
Magnets | Liquid Level Instrumentation
Technical Support | Customer
Services | AMI Picturebook | E-Mail AMI