| Program Out Voltage |
|
| Programmable Limits: |
-10 to +10 VDC (voltage-voltage mode) |
| Accuracy: |
0.1% of Vmax |
| Temperature Coefficient: |
0.005% of Vmax / °C |
| Resolution: |
20 µV |
| Stability: |
< 35 mV P-P when paused or holding |
| Magnet Voltage Measurement |
|
| Maximum Limits: |
-20 to +20 VDC |
| Accuracy: |
0.1% of Vmax |
| Temperature Coefficient: |
0.01% of Vmax / °C |
| Resolution: |
10 mV |
| Persistent Switch Heater Output |
|
| Programmable Limits: |
0.1 to 100 mA DC |
| Accuracy: |
0.5 mA |
| Temperature Coefficient: |
0.02 mA / °C |
| Maximum Compliance: |
13.5 V |
| Resolution: |
0.1 mA |
| Optional External Supply
Limits: |
10 VA, 0.5 A max, 100 VDC max |
| Power Requirements |
|
| Primary: |
100-120 or 200-240 VAC ±10%
50-60 Hz, 50 VA max |
| Memory Backup Battery: |
3.6 Volt AA Lithium Cell |
| Physical |
|
| Dimensions: |
88 mm H x 483 mm W x 191 mm D
(3.5" H x 19" W x 10.75" D) |
| Weight: |
4.2 kg (9.2 lbs). |
| Torque Limits
on Current Shunt Terminals: |
5, 10, and 100 A models: 50 in-lbs.
200 A model: 150 in-lbs.
300 A model: 360 in-lbs. |
| Environmental |
|
| Ambient Temperature: |
Operating: 0 °C to 50 °C (32 °F to 122 °F)
Non-operating: -20 °C to 60 °C (-4 °F to 140 °F) |
| Relative Humidity: |
80% up to 31 °C (88 °F), decreasing
linearly to 50% at 50 °C (122 °F) |
| Altitude: |
2000 m (6562 ft.) Indoor use |
| Standards |
|
| EMI/EMC Standards: |
EN50082-1
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN55022, Class A |
| Safety Standard: |
EN61010-1 |
| Installation Category: |
Pollution Degree 2, Overvoltage Category II
as defined by IEC664 |
| a Resolution of the
IEEE 754 double-precision floating point type consisting of a 52-bit
fraction and 11-bit exponent. |
Please contact an AMI Sales Representative
for more detailed information including compatible power supplies, function descriptions,
communication command set, system diagrams, etc.