Model 186 Liquid Level Controller

Product Brochure

The American Magnetics, Inc. (AMI) Model 186 Liquid Level Controller (cryogen monitor) system is an advanced, microprocessor-based solution designed to provide accurate and reliable level monitoring and control of virtually any cryogenic liquid.

Capacitance-based level sensing                  

Simple calibration

Automated fill and alarm functions

The Model 186 fill control and alarm functions are front panel programmable from 0 to 100 percent of the active sensing length. Two independent control band setpoints, "A" and "B", activate a power receptacle on the rear panel. When the liquid level drops below the "B" setpoint, the power receptacle is energized and remains energized until the liquid level rises to the "A" setpoint. The Model 186 also provides a fail-safe timer feature that automatically de-energizes the power receptacle once a user programmed maximum time interval of up to 600 minutes is reached. The Model 186 is ideal for unattended systems where automated fill is required.

In addition, the Model 186 provides "High" and "Low" alarm setpoints to activate front panel LED warning indicators and rear panel relay outputs in the event of an overfill or liquid loss condition. The "Low" level alarm also energizes an audible warning which can be silenced from the front panel.

Convenient display

The instrument is equipped with a 4-digit LED display which provides liquid level indication in inches, centimeters, or percent as selected by a front panel switch. A front panel switch allows the user to easily adjust the instrument's length setting for a specific active sensor length. The sensor active length can be entered in either inches or centimeters.

Remote monitoring or control

The Model 186 can be provided with an optional 0-10 volt DC signal on the rear panel of the instrument for use with a recorder. A 4-20 mA current loop option is available in lieu of the voltage signal. Computer interface options, including RS-232 Serial Port/Data Logger or IEEE-488, are also available.

Model 185 Specifications at 25 °C

Level Measurements
Resolution: 0.1%, 0.1 cm, or 0.1 in
Linearity: ± 0.1 %
Maximum Length Readout: 650.0 cm (255.9 in)
Operating Parameters
HI and LO Alarms: 0% to 100% adjustable
HI/LO Alarm Relay Contact Ratings: 30 VAC or 60 VDC, 10 VA (up to 0.5 A max)
[normally open, closed on alarm]
A and B Control Setpoints: 0% to 100% adjustable
Controller Output: AC line voltage @ 2A max current
Fill Timer: 0.1 to 600.0 minutes
0-10 Volt Analog Output
Integral Non-linearity: ± 0.012 %
Resolution: 16 bits
Total Error: ± 1.1 %
Voltage Drift (0-10 V): 100 ppm  / °C
Maximum Load: 4 mA (2.5 kW @ 10 V output)
4-20 mA Analog Output @ 24 V
Vext Supply Range: 13-32 VDC
Voltage Compliance: Vext - 3.5
Integral Non-linearity: ± 0.012 %
Resolution: 16 bits
Total Error: ± 0.25 %
Current Drift (4-20 mA): 75 ppm  / °C
PSRR: 10 mA / V
Power Requirements
Primary: 100-120 or 200-240 VAC ±10% 50-60 Hz, 
2.2 A
Physical
Dimensions (Standard): 97 mm H x 213 mm W x 282 mm D
(3.8" H x 8.4" W x 11.1" D)
Weight (Standard): 1.6 kg (3.6 lbs)
Dimensions (Rack Mount): 89 mm H x 483 mm W x 282 mm D
(3.5" H x 19" W x 11.1" D)
Weight (Rack Mount): 2.0 kg (4.3 lbs)
Environmental
Ambient Temperature: Operating: 0 °C to 50 °C (32 °F to 122 °F)
Non-operating: -20 °C to 60 °C (-4 °F to 140 °F)
Relative Humidity: 0% to 95%; non-condensing

ce.gif (892 bytes)

Microprocessor-based electronics provide 0.1% readout accuracy. Nonvolatile, read only memory maintains instrument calibration without battery backup. Watchdog timer circuitry and low line voltage (brownout) detector prevent microprocessor lock-up and provide fail-safe operation.